The cause of the low sulfur analyzer
The cause of the low sulfur analyzer
The reasons for the low results of the Kulun survey
Measuring sulphur instrument low results directly affect the measurement accuracy, even causing air pollution, so why measuring sulphur instrument will result in low?, the following by Hongke coal testing equipment for you a simple analysis, resulting in measuring sulphur instrument results and the reason of the low what?
One, the measuring instrument of the different diameter pipe (quartz tube) fracture or no plug in the end, leading to gas road sealing is not strict, resulting in low sulfur meter.
Two, the gas flow is too low, which affects the flow of gas, resulting in the low and unstable.
1, air flow meter is not adjusted in place, you can adjust the air flow.
2, electrolysis pool in black, Dusseldorf flow. Can change the fuse plate.
3, the air purifier in the pump power is reduced, the replacement of aging, air pump.
Three, the electrolyte failure, resulting in low or unstable test results of the test instrument.
Four, the electrolytic cell pole piece is dirty, has the detail of the tiny yellow adhesion layer, can use the blade to blow on the pole piece to solve.
Five, Kulun sulfur analyzer / sulfur analyzer, although the results are low, but the parallel sample is very good, you can adjust the accuracy of the measuring instrument to solve the problem of low.
Note: when the results of the test instrument is low, do not immediately adjust the accuracy of the instrument to measure the sulfur meter, must first rule out the above five points outside the reasons, and then adjust the accuracy.
With people's life continuously improving, we pay more and more attention to the natural environment, air quality, in the new energy development and cost control, coal is still the main power fuel, coal sulfur content, directly affect the atmosphere. It is imperative to detect the total sulfur content of coal, and the Kulun sulfur analyzer / sulfur analyzer is an important instrument for the detection of total sulfur in coal, and has been widely used.
Measuring sulphur instrument low results directly affect the measurement accuracy, even causing air pollution, so why measuring sulphur instrument will result in low?, the following by Hongke coal testing equipment for you a simple analysis, resulting in measuring sulphur instrument results and the reason of the low what?
One, the measuring instrument of the different diameter pipe (quartz tube) fracture or no plug in the end, leading to gas road sealing is not strict, resulting in low sulfur meter.
Two, the gas flow is too low, which affects the flow of gas, resulting in the low and unstable.
1, air flow meter is not adjusted in place, you can adjust the air flow.
2, electrolysis pool in black, Dusseldorf flow. Can change the fuse plate.
3, the air purifier in the pump power is reduced, the replacement of aging, air pump.
Three, the electrolyte failure, resulting in low or unstable test results of the test instrument.
Four, the electrolytic cell pole piece is dirty, has the detail of the tiny yellow adhesion layer, can use the blade to blow on the pole piece to solve.
Five, Kulun sulfur analyzer / sulfur analyzer, although the results are low, but the parallel sample is very good, you can adjust the accuracy of the measuring instrument to solve the problem of low.
Note: when the results of the test instrument is low, do not immediately adjust the accuracy of the instrument to measure the sulfur meter, must first rule out the above five points outside the reasons, and then adjust the accuracy.
With people's life continuously improving, we pay more and more attention to the natural environment, air quality, in the new energy development and cost control, coal is still the main power fuel, coal sulfur content, directly affect the atmosphere. It is imperative to detect the total sulfur content of coal, and the Kulun sulfur analyzer / sulfur analyzer is an important instrument for the detection of total sulfur in coal, and has been widely used.